Olso
Metal Ions Analyzer
APPLIED ANALYTICSProcess Analyzer

Metal Ions Analyzer

AAI’s metal ion analyzer continuously draws a liquid sample and outputs real-time concentrations of metal ions including copper, iron, nickel, chromium, and cobalt.

Measurement PrincipleDispersive ultraviolet-visible (UV-Vis) absorbance spectrophotometry
Detectornova II™ UV-Vis diode array spectrophotometer
Spectral Range200–800 nm
Light SourcePulsed xenon lamp (average 5 year lifespan)
Signal Transmissio600 μm core 1.8 meter fiber optic cables. Other lengths available
Path LengthApplication-dependent
Sample ConditioningCustom design per application
Analyzer CalibrationIf possible, analyzer is factory calibrated with certified calibration fluids; no re-calibration required after initial calibration; measurement normalized by Auto Zero
Reading VerificationSimple verification with samples and self-check diagnostic
Human Machine InterfaceIndustrial controller with touch-screen LCD display running ECLIPSE™ Software
Data StorageSolid State Drive
OPERATING CONDITIONS
Analyzer EnvironmentIndoor/Outdoor (no shelter required)
Ambient TemperatureStandard: 0 to 35 °C (32 to 95 °F), Optional: -20 to 55 °C (-4 to 131 °F), To avoid radiational heating, use of a sunshade is recommended for systems installed in direct sunlight.
Sample TemperatureStandard: -20 to 70 °C (-4 to 158 °F), Optional: up to 150 °C (302 °F) with cooling extensions, Contact AAI for temperatures above 150 °C (302°F)
Sample PressureUsing standard flow cell: 206 bar (3000 psi)
UTILITIES
Electrical85–264 VAC 47–63 Hz
Power Consumption45 watts
OUTPUTS
Standard Outputs1x galvanically isolated 4–20mA analog output per measured analyte (up to 3; additional available by upgrade), 2x digital outputs for fault and SCS control
Optional OutputsModbus TCP/IP; RS-232; RS-485; Fieldbus; Profibus; HART
PHYSICAL SPECIFICATIONS
Analyzer TypesOMA-300 Wall-Mounted Analyzer, OMA-206P Portable Analyzer, OMA-406R Rackmount Analyzer
How can we help you?